28. ATOMIC EMISSION SPECTROSCOPY

Atomic Emission Spectroscopy (AES) is an analytical technique used to determine the elemental composition of a sample by measuring the light emitted by atoms excited at high temperatures. When atoms are supplied with sufficient energy, their electrons jump to higher energy levels. As they return to the ground state, they release energy in the form of light. This emission is characteristic for each element, allowing both qualitative and quantitative analysis.

AES is widely used in pharmaceuticals, metallurgy, environmental testing, petrochemicals, and clinical laboratories due to its sensitivity, speed, and ability to analyze multiple elements simultaneously.


Principle

The principle of Atomic Emission Spectroscopy is based on the excitation of atoms by thermal or electrical energy. Once excited, these atoms emit light at wavelengths unique to each element. The emitted light is resolved by a monochromator or spectrograph and is detected using a photosensitive detector.

The emission intensity is directly proportional to the concentration of the element in the sample.


Instrumentation

AES instrumentation consists of several major components responsible for excitation, wavelength selection, and detection of emitted radiation.

Spectroscopic Sources

Sources provide the energy required to transform analyte atoms into excited states. Common sources include:

  • Flame Sources: Use air–acetylene or nitrous oxide–acetylene flames to excite easily ionizable metals.
  • Plasma Sources (ICP): Inductively Coupled Plasma produces extremely high temperatures suitable for multi-element analysis with high accuracy.
  • DC Arc and AC Spark Sources: Useful for solid samples like metals and alloys.
  • Microwave Plasma: Low-power plasmas used for environmental and routine sample testing.

Sample Holder

The sample holder varies depending on the sample form—solid, liquid, or solution. For solutions, a nebulizer converts the liquid into fine aerosol droplets which are introduced into the excitation source. For solid metals, direct arc or spark excitation is used.


Monochromator

A monochromator isolates specific wavelengths emitted by the excited atoms. It typically includes:

  • Entrance slit to admit emitted radiation.
  • Dispersive element such as a prism or diffraction grating.
  • Exit slit to select the required wavelength.

Grating monochromators offer high resolution and are widely used in modern AES instruments.


Detector

Detectors convert the emitted light into electrical signals. Several types of photoelectric detectors are used:

1) Photodiodes

Photodiodes convert light into electrical current using the photoelectric effect. They offer fast response, compact size, and good linearity but are less sensitive than PMTs for very low-light measurements.

2) Photomultiplier Tubes (PMTs)

PMTs amplify weak light signals through secondary electron emission across multiple dynodes. They are highly sensitive and ideal for detecting trace metal emissions. However, they require high-voltage supplies and are sensitive to intense light.

3) Charge-Coupled Devices (CCDs)

CCDs capture light across thousands of pixels, allowing simultaneous detection of multiple wavelengths. They offer excellent spatial resolution and broad wavelength sensitivity but may require cooling to reduce noise.

4) Avalanche Photodiodes (APDs)

APDs operate under high reverse-bias voltage, providing internal signal amplification. They offer higher gain than photodiodes but produce more noise than PMTs.


Spectrograph

A spectrograph disperses emitted light over a range of wavelengths and projects it onto a detector array. Spectrographs allow simultaneous multi-element detection and are commonly used in ICP-AES systems for high-throughput analysis.


Interferences in Atomic Emission Spectroscopy

Several interferences can affect accuracy:

  • Chemical Interference: Formation of refractory compounds may reduce emission.
  • Spectral Interference: Overlapping emission lines from other elements.
  • Background Emission: Flame, plasma, or matrix components may emit light that interferes with signal.
  • Physical Interference: Variations in viscosity, surface tension, and sample introduction.

These can be minimized using internal standards, background correction, and optimized operating conditions.


Applications

Atomic Emission Spectroscopy is widely applied in multiple fields due to its sensitivity and ability to detect several elements simultaneously:

  • Determination of sodium, potassium, lithium, calcium, and magnesium in pharmaceuticals.
  • Trace metal analysis in biological fluids such as blood or urine.
  • Environmental monitoring of water, air, and soil pollutants.
  • Quality control of metals, alloys, and industrial samples.
  • Petrochemical analysis for metal contaminants.
  • Food quality testing for mineral content.

Advantages of Atomic Emission Spectroscopy

  • Rapid multi-element analysis.
  • High sensitivity for alkali and alkaline earth metals.
  • Suitable for trace and ultra-trace detection.
  • Minimal sample preparation in ICP-AES.
  • Applicable to solids, liquids, and gases.

Disadvantages of Atomic Emission Spectroscopy

  • Instrumentation can be expensive, especially ICP-based systems.
  • Spectral interference may complicate analysis.
  • Requires high skill for method optimization.
  • Not ideal for elements with low excitation efficiency in flames.

Additional Information: Photoelectric Detection in AES

Photoelectric detection plays a central role in AES because it enables conversion of emitted light into measurable electrical signals. The detector chosen influences sensitivity, linearity, response speed, and spectral range. PMTs remain the most widely used detectors due to their exceptional sensitivity, while CCDs are preferred in modern multi-element spectrographs.

Detailed Notes:

For PDF style full-color notes, open the complete study material below: